N9927-90003
3-7
FieldFox Analyzer
Tests and Adjustments
Operational Check
Operational Check
This operational check exercises only very basic measurement functionality.
Required Equipment
A short compatible with the port connector type on the FieldFox.
A load compatible with the port connector type on the FieldFox.
(Both are typically available in a compatible cal kit.)
Procedure
1. Press
Preset > Preset (Factory)
. The top-left corner of the FieldFox display indicates the active
measurement type following the preset.
2. If the active measurement is SA then skip to step 5 below.
If the active measurement is S11 or CAT then complete this step below:
•
Connect a short to Port 1.
•
Displayed trace should be within
±
4 dB of zero.
•
Connect a load to Port 1.
•
Displayed trace should approximate the value of load, typically it will be below
−
20 dB (20 dB
for CAT mode).
3. Press
Mode
. If NA is not available, skip to step 5 below.
If NA is available then continue at step 4 below.
4. Press
NA
then press
Measure (1)
. If S22 is not available then skip to step 5 below.
If S22 is available then continue this step below:
•
Connect a short to Port 2.
•
Press
S22
•
Displayed trace should be within
±
4 dB of zero.
•
Connect a load to Port 2
•
Displayed trace should approximate the value of load, typically it will be below
−
20 dB.
Содержание FieldFox MW N9913A
Страница 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Страница 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Страница 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Страница 17: ...N9927 90003 2 1 2 General Product Information ...
Страница 26: ...2 10 N9927 90003 General Product Information FieldFox Analyzer Agilent Support Services and Assistance ...
Страница 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Страница 37: ...N9927 90003 4 1 4 Troubleshooting ...
Страница 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Страница 51: ...N9927 90003 5 1 5 Theory of Operation ...
Страница 58: ...5 8 N9927 90003 Theory of Operation FieldFox Analyzer Measurement Group Figure 5 2 Simplified RF Block Diagram ...
Страница 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Страница 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Страница 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Страница 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Страница 149: ...N9927 90003 Appendix 1 Appendix A ...
Страница 156: ...Index 4 N9927 90003 Index ...