Repair and Replacement Procedures
Removing and Replacing the Measurement Group Assemblies
Replacing the RF Top Panel Assembly
1. Separate the front and rear cases. Refer to
“Separating the Front and Rear Cases” on page 7-10
.
2. Remove all measurement group assemblies as one unit. Refer to
Group Assemblies as One Unit” on page 7-27
3. Refer to
. Unplug the two connector plugs (tethered caps) from the type-N RF
connectors.
4. Disconnect cable W3 and W4.
5. Remove the six screws (item 1) and lift the RF top panel over the type-N RF connectors.
6. Position the new RF top panel over the type-N RF connectors and loosely reinstall the six screws
(item 1). Do not tighten these screws at this time; they will be tightened after the front and rear cases
have been combined.
7. Connect cable W3 and W4 to the A5 System board connectors. There should be an audible click
when the connector is fully in place.
8. Position (tuck) the cables W3 and W4 into the cavity above the A5 System board to avoid
interference with nearby parts.
9. Plug the two connector plugs (tethered caps) back into the type-N RF connectors to protect the
connectors.
10. Perform the steps under
“Post-replacement Procedure” on page 7-44
Figure 7-19
Replacing the RF Top Panel Assembly
Содержание FieldFox MW N9913A
Страница 8: ...Contents 4 Service Guide N9927 90003 Contents ...
Страница 9: ...N9927 90003 1 1 1 Safety and Regulatory Information ...
Страница 16: ...1 8 N9927 90003 Safety and Regulatory Information FieldFox Analyzer Regulatory Information ...
Страница 17: ...N9927 90003 2 1 2 General Product Information ...
Страница 26: ...2 10 N9927 90003 General Product Information FieldFox Analyzer Agilent Support Services and Assistance ...
Страница 27: ...N9927 90003 3 1 3 Tests and Adjustments ...
Страница 37: ...N9927 90003 4 1 4 Troubleshooting ...
Страница 50: ...4 14 N9927 90003 Troubleshooting FieldFox Analyzer Measurement Group Troubleshooting ...
Страница 51: ...N9927 90003 5 1 5 Theory of Operation ...
Страница 58: ...5 8 N9927 90003 Theory of Operation FieldFox Analyzer Measurement Group Figure 5 2 Simplified RF Block Diagram ...
Страница 81: ...N9927 90003 6 1 6 Replaceable Parts ...
Страница 95: ...N9927 90003 7 1 7 Repair and Replacement Procedures ...
Страница 140: ...7 46 N9927 90003 Repair and Replacement Procedures FieldFox Analyzer Post Repair Procedures ...
Страница 141: ...N9927 90003 8 1 8 General Purpose Maintenance and Service Procedures ...
Страница 149: ...N9927 90003 Appendix 1 Appendix A ...
Страница 156: ...Index 4 N9927 90003 Index ...