Real-Time I/Q Baseband Generator User’s and Programming Guide
2-31
HP ESG Family Signal Generators
Using Functions
Options UN8 and 202
Creating User-Defined FIR Filters
Modifying the Coefficients
1. The value for coefficient 0 should be highlighted. Use the front panel knob to scroll
down until coefficient 15 is highlighted.
2. Press
0
>
Enter
to change the value of the coefficient to 0.
3. Press
Display FFT
to see the effects of the change:
Notice that the graphic display can provide a useful troubleshooting tool (in this case
indicating a missing coefficient value for a proper Gaussian response).
4. To return to the menu keys, press
Return
.
5. In addition to changing existing values, you can also insert and delete rows of
coefficients and change the oversample ratio. Press
More (2 of 2)
to access these softkeys.