Real-Time I/Q Baseband Generator User’s and Programming Guide
4-35
ESG Family Signal Generators
Programming Commands and Examples
Options UN8 and 202
Programming Commands
RESet
In Reset & Run External Frame Trigger mode, the trigger input latch is
reset by a register bit in the status byte. When the register bit is released,
the next trigger signal will reset and start the pattern. Prior to the trigger
signal, the pattern waits at last bit until triggered. Pattern will continue to
run repeatedly, independent of trigger until trigger input latch bit is
toggled in the status byte. Once the trigger input latch register bit is
toggled, the pattern will play to the end and wait at the last bit for the
next trigger signal.
MANual
In Manual Reset & Run External Frame Trigger mode, pattern is reset
immediately (without running to the end) and restarts and runs
repeatedly until next trigger transition to active.
*RST value: Single Shot
Frame Repeat
Custom
[:SOURce]:RADio:CUSTom:REPeat SINGle|CONTinuous
[:SOURce]:RADio:CUSTom:REPeat?
Other Formats
[:SOURce]:RADio:<desired format>:REPeat SINGle|CONTinuous
[:SOURce]:RADio:<desired format>:REPeat?
This command sets the output of the selected frame:
SINGle
Outputs one occurrence of the selected frame.
CONTinuous
Outputs a continuous stream of the selected frame.
*RST value: Continuous
Frame Transmission Rate
[:SOURce]:RADio[:NADC]:FRATe FULL|HALF
[:SOURce]:RADio[:NADC]:FRATe?
[:SOURce]:RADio:PDC:FRATe FULL|HALF
[:SOURce]:RADio:PDC:FRATe?
This command sets the transmission rate for your framed data pattern.
For NADC format, the choices are Full (timeslots 1, 2, and 3 are paired with timeslots 4, 5,
and 6, respectively), or Half (6 individual timeslots).
For PDC format, the choices are Full (timeslots 0, 1, and 2 are paired with timeslots 3, 4,
and 5, respectively), or Half (6 individual timeslots and faster transmissions).
Notice that your selection, either full-rate or half-rate, is shown on the display directly
above the timeslot pattern visual representation.
*RST value: Full