Chapter 1
83
Instrument Messages
Instrument Messages
In
str
u
me
n
t Me
s
s
a
g
es
10662
Inconsistent sync word
The instrument detected the training sequence, but the training sequence
number is inconsistent between the AM path and the PM path.
10701 to 10800: W-CDMA Error Messages (Option BAF)
An error number in this
range indicates the instrument has detected an error relating to the W-CDMA personality.
10702
Signal too noisy
10704
Input power too low
The instrument only identifies a channel as active if it meets the default
threshold criteria that it is within 20 dB of the highest power code
channel. The threshold can be changed using the Active Set Threshold
function in the Meas Setup menu.
10768
Cannot correlate to input signal
Cannot correlate to the input signal and no active channel is found. (from
the composite EVM measurement) An active channel must meet the
default threshold criteria that it is within 20 dB of the highest power code
channel. The threshold can be changed using the active set threshold
function in the Meas Setup menu.
10772
Burst not found.
Either the signal being analyzed has insufficient power, the rising or
falling edges cannot be detected, or the burst is less than
126 microseconds.
10776 PSA
Cannot sync DPCCH pilot.
Cannot synchronize measurement with DPCCH pilot for Power Control
measurement, because the pilot signal cannot be found. Make sure
DPCCH is present in the W-CDMA uplink signal, and that the slot format
and scramble code are set correctly.
10778 PSA
Valid PRACH burst not found.
Cannot synchronize measurement with PRACH channel for Power Control
measurement, because the signal cannot be found. Make sure PRACH is
present in the W-CDMA uplink signal, and that the preamble signature
and scramble code are set correctly.
10780
Critical memory shortage occurred.
This error message appears when a critical memory shortage occurs, and
the current measurement is aborted. To continue, restart the
measurement by pressing the
Preset
key, or by exiting and then
re-entering the current mode or the current measurement, or cycle the
power.
Содержание ESA Series
Страница 6: ...6 Contents Table of Contents ...
Страница 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Страница 33: ...33 Instrument Messages 1 Instrument Messages ...
Страница 111: ...111 Functional Tests 2 Functional Tests ...
Страница 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Страница 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Страница 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...