10524 to 10561: Phase Noise Error Messages (Option 226)
An error number in this
range indicates the instrument has detected an error relating to the Phase Noise
personality.
10524
This measurement does not support the *.CSV file format.
You cannot load or save base instrument traces, as this is not supported by
the Log Plot measurement.
10525
Invalid format: CSV files are not supported by this
measurement.
You cannot load or save base instrument traces, as this is not supported by
the Log Plot measurement.
10525
Use View/Trace menu when loading or saving logarithmic traces.
You cannot load CSV format trace files or save traces in the CSV format in
the Log Plot measurement of the Phase Noise Measurement Personality.
10526
Cannot display Ref Trace because it has no data.
Reference trace cannot be displayed, as there is currently no data assigned
to it - use the functions under the Store Ref Trace menu, or load a trace to
assign some data.
10527
Invalid Marker Trace. Marker moved to Raw Trace.
Cannot place markers on the reference trace, because the reference trace
is currently turned off or has no data.
10529
Failed to Load trace. Bad file format.
The load trace operation could not be completed, as the input file was not
in the expected format. You can only load traces that were previously
saved using the 'Save Trace' feature.
10530
Trace file contains no compatible traces.
The trace file may have been created by another version of the Phase Noise
personality, which uses a different trace format that is incompatible with
the version you are running. Please check you are running the most up to
date version of the personality.
10531
Trace file was created by incompatible version of Opt. 226
The trace file may have been created by another version of the Phase Noise
personality, which uses a different trace format that is incompatible with
the version you are running. Please check you are running the most up to
date version of the personality.
10532
Cannot open trace file for writing. Save Failed.
Cannot write the trace file to the destination filename. This could be
because the disk is full, or possibly due to a filename error. If using a
floppy disk, check there is a formatted disk in the drive.
Содержание ESA Series
Страница 6: ...6 Contents Table of Contents ...
Страница 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Страница 33: ...33 Instrument Messages 1 Instrument Messages ...
Страница 111: ...111 Functional Tests 2 Functional Tests ...
Страница 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Страница 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Страница 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...