Chapter 1
67
Instrument Messages
Instrument Messages
In
str
u
me
n
t Me
s
s
a
g
es
10182
Lower Custom Mask is Invalid!
The user-specified lower custom mask cannot be resolved into a limit line.
The format may be bad.
10183
Resolution BW
<
300kHz.
The test will not meet EDGE specifications because the resolution BW has
fallen below 300 kHz.
10184
Cannot find the Power vs Time Limits File.
Cannot find the limit line definition file for the EDGE standards. This
message is displayed while the
Measure
key is grayed out.
10185
Gate option not installed. Results may not be accurate.
10186
Measurement does not support the current radio standard.
The measurement you have chosen is currently greyed out. Select a radio
standard which is supported or configure this measurement manually
(
Mode Setup
,
Radio Std
.
None
).
10186
EDGE EVM only supports EDGE TCH burst type (Burst Type)
10187
Radio standard is not supported by the current measurement.
The standard you have chosen is currently greyed out. Select a radio
standard which is supported by the current measurement or configure this
measurement manually (
Mode Setup
,
Radio Std
.
None
).
10187
EDGE ORFS only supports EDGE TCH burst type (Burst Type)
10188
EDGE PVT only supports EDGE TCH burst type (Burst Type)
10189
EDGE PVT only supports EDGE TCH burst type (Ref Burst)
10190
Select a valid Radio Standard/Measurement combination to
enable.
The Enable All Measurements menu key was pressed when it was
disabled. To enable select a radio standard that is supported by the current
measurement or a measurement that supports the current radio standard.
10190
EDGE ORFS only supports EDGE TCH burst type (Ref Burst)
10191
Mask unavailable for current Span - increase to display mask.
The current span setting is either narrower than the mask width or so
wide that there are too few display points to allow the mask to be drawn.
Increase or decrease the span to display the mask.
10191
EDGE EVM only supports EDGE TCH burst type (Ref Burst)
Содержание ESA Series
Страница 6: ...6 Contents Table of Contents ...
Страница 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Страница 33: ...33 Instrument Messages 1 Instrument Messages ...
Страница 111: ...111 Functional Tests 2 Functional Tests ...
Страница 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Страница 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Страница 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...