
64
Chapter 1
Instrument Messages
Instrument Messages
Ins
tr
u
me
nt Messa
g
e
s
10148
B7D DSP Code Fail. Reload Opt BAH. Contact your Service
Center.
The Digital Signal Processor algorithm files required to perform the
demodulation are not present in the IDSP: drive.
10149
Opt BAH DSP algorithm code file requires upgrade.
The Digital Signal Processing algorithm code file revision is not supported
by the currently loaded personality version. Refer to the user guide for the
personality in use for more information on installation/upgrade.
10150
Opt BAH DSP algorithm coef. file requires upgrade.
The Digital Signal Processor algorithm coefficient file revision is not
supported by the currently loaded personality version. Refer to the user
guide for the personality in use.
10151
Opt BAH DSP algorithm files failed to load, aborting measure.
The Digital Signal Processor algorithm files required to perform the
demodulation are corrupt and cannot be loaded properly. Reinstall
measurement personality.
10151
B7D DSP Code Fail. Reload Opt BAH. Contact your Service
Center
.
The Digital Signal Processor algorithm files required to perform the
demodulation are present in the IDSP: drive and cannot be loaded
properly into DSP RAM. Reinstall measurement personality.
10152
Lost trigger, aborting measurement.
The selected trigger source was present at the start of the measurement,
but timed out before the measurement completed.
10153
DSP algorithm timeout, aborting measurement
The Digital Signal Processor demodulation algorithm timed-out for an
unknown reason. This message normally indicates a problem with the
modulated signal.
10154
Measurement not defined for Out of Band.
You have changed to an out-of-band frequency range. The band
measurement only operates in the selected band.
10155
No Fast ADC hardware installed
. Meas unavailable.
The analyzer cannot use sweep times of less than 5 msec when Option B7D
or Option AYX is not installed. Therefore, the measurement will not be
executed.
Содержание ESA Series
Страница 6: ...6 Contents Table of Contents ...
Страница 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Страница 33: ...33 Instrument Messages 1 Instrument Messages ...
Страница 111: ...111 Functional Tests 2 Functional Tests ...
Страница 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Страница 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Страница 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...