![Agilent Technologies ESA Series Скачать руководство пользователя страница 62](http://html1.mh-extra.com/html/agilent-technologies/esa-series/esa-series_manual_2868109062.webp)
62
Chapter 1
Instrument Messages
Instrument Messages
Ins
tr
u
me
nt Messa
g
e
s
Greater than 10000: Measurement Applications Errors
NOTE
Error numbers are displayed in the error queue,
not
on the display.
To see an error
number
, press
System
,
Show Errors
.
10008 to 10522: Power Suite Error Messages
An error number in this range indicates
the instrument has detected an error relating to the Power Suite functionality.
10008
Preferred resolution bandwidth not available.
The calculated required resolution bandwidth for this measurement is not
available.
10010
One or more harmonics past freq limit: number decreased
.
Highest harmonic was past the frequency limit of the analyzer, so the
number of measured harmonics was decreased.
10011
First harmonic is past analyzer frequency limit.
First harmonic was past the frequency limit, so no harmonics can be
measured. The analyzer will simply measure the fundamental. If you have
an Option AYZ it allows the use of Agilent 11970 Series, and Agilent 11974
external mixers to extend the frequency range to 110 GHz. Operation to
325 GHz is also possible with non-Agilent mixers.
10012
No Fundamental > 0 Hz found in given span.
There were no frequencies greater than 0 Hz in the starting span, so the
centre frequency at measurement entry will be used as the fundamental.
10013
No Fundamental > -50 dBm found in given span.
There was no fundamental found, so the centre frequency at measurement
entry will be used as the fundamental.
10020
Signal Tracking is not available when Noise Correction is on
.
If signal tracking is turned on while noise correction is on, it will be turned
off. To use the signal tracking function, first turn noise correction off.
10138
Valid GSM burst not found in frame (Burst Type)
No active GSM bursts that match the selected burst type have been
detected in the RF input signal. Search was performed over a complete
GSM frame.
10139
Valid GSM burst not found in specified timeslot (Burst Type)
No active GSM bursts that match the selected burst type have been
detected in the RF input signal. Search was performed over the specified
timeslot.
Содержание ESA Series
Страница 6: ...6 Contents Table of Contents ...
Страница 32: ...32 Instrument Messages Alphabetical Listing Instrument Messages ...
Страница 33: ...33 Instrument Messages 1 Instrument Messages ...
Страница 111: ...111 Functional Tests 2 Functional Tests ...
Страница 118: ...118 Chapter2 Functional Tests Test Equipment Functional Tests ...
Страница 119: ...119 PSA Functional Tests 3 PSA Functional Tests ...
Страница 139: ...139 ESA Functional Tests 4 ESA Functional Tests ...