Installing the Tinted Acrylic Door

Installation Note

!

"

Installation Note

Installing the Tinted Acrylic Door

This procedure describes how to install the Tinted Acrylic Door on the 
Agilent E840xA mainframe. The door requires mainframe Standard Adapter 
Kit Option AXA (p/n E8400-80923), and is compatible with either the 
support rail kit (Agilent E3664AC) or the rack slide kit (p/n 1494-0411).

Parts List

The parts included in the door kit are shown in Table 1.

Note

The door requires the mainframe to be recess mounted a minimum of 
111 mm (4.36") or up to a maximum of 270.7 mm (10.6 inches) to 
accommodate wiring terminals on the VXI instruments. When using the 
minimum recess distance, field wiring cables must enter/exit the 
instruments through the top and/or bottom openings only.

Procedure

1. Recess mount the mainframe a minimum of 111 mm (4.36 inches) 

using the Standard Adapter Kit and either the support rail kit or rack 
slide kit (see E8400-90038 for mainframe mounting instructions). 

Note

For the minimum recess distance, attach the rack mount adapters to the 
mainframe beginning with the 12th hole from the front of the adapter.

2. If necessary, remove the dress screws (0570-1577) securing the 

adapters to the rack. 

3. Attach the hinge pins to the right rack mount adapter using four 10-32 

x 0.375 flat head screws (Figure 1). Attach the rubber door stops and 
door latch to the left adapter using four m3x8 pan head screws as also 
shown in Figure 1.

Table  1.  Agilent E840xA Door Kit Parts List.

Quantity

Description

Part Number

1

Acrylic Door Assembly

E8400-09301

2

Hinges

3110-0409

1

Door Latch, Lock, and Keys

1390-0891

2

Rubber Door Stops

0403-0002

4

Screws - 10-32x0.375 flat head

2680-0100

4

Screws - m3x8 pan head

0515-0897

*E8400-90039*

Manual Part Number: E8400-90039

                     Edition 3   E0309