Error Term Data
Source Match (E
SF
and E
SR
)
E
SF
and E
SR
are the forward and reverse uncorrected source match terms of the driven
port. They are obtained by measuring the reflection (S
11
, S
22
) of an open, and a short that
are connected directly to the ports. Source match is a measure of the match of the coupler,
as well as the match between all components from the source to the output port.
The measurements most affected by source match errors are reflection and transmission
measurements of highly reflective DUTs.
Typical Cause of Failure
The
calibration kit open or shor
t is the most common cause of source match specification
failure.
If the open or short performance has been independently verified, then suspect the
analyzer
switch splitter, step attenuator, or coupler
.
To troubleshoot, refer to
“Checking the Signal Separation Group” on page 4-46
.
Figure A-4 Typical E
SF
/E
SR
Without Cables
Содержание E8362A
Страница 11: ...Service Guide E8364 90001 1 1 1 Safety and Regulatory Information ...
Страница 19: ...Service Guide E8364 90001 2 1 2 General Product Information ...
Страница 29: ...Service Guide E8364 90001 3 1 3 Tests and Adjustments ...
Страница 79: ...Service Guide E8364 90001 4 1 4 Troubleshooting ...
Страница 139: ...Service Guide E8364 90001 5 1 5 Theory of Operation ...
Страница 169: ...Service Guide E8364 90001 6 1 6 Replaceable Parts ...
Страница 215: ...Service Guide E8364 90001 7 1 7 Repair and Replacement Procedures ...
Страница 287: ...Service Guide E8364 90001 A 1 A Error Terms ...
Страница 302: ...A 16 Service Guide E8364 90001 Error Terms PNA Series Microwave Network Analyzers Error Term Data E8362A E8363A E8364A ...
Страница 303: ...Service Guide E8364 90001 B 1 B Option Enable Utility ...
Страница 309: ...Service Guide E8364 90001 C 1 C Firmware Upgrades ...
Страница 313: ...Service Guide E8364 90001 D 1 D Operating System Recovery ...