Chapter 5
239
Language Reference
DISPlay Subsystem
is not set above the reference level and the reference is set so
that the signal is within the display range. When in fixed mode
the measurement has approximately 70 dB of display range.
Front Panel
Access:
AMPLITUDE/Y Scale, IF Gain Auto Fixed
(front panel access
is available with firmware revision A.06.00 and later).
Normalized Reference Level
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel <rel_ampl>
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRLevel?
Sets the normalized reference level.
NOTE
See command
:CALCulate:NTData[STATe] OFF|ON|0|1
Factory Preset
and *RST:
0 dB
Range:
–327.6 to 327.6 dB
Default Unit:
Current active units
Front Panel
Access:
View/Trace, Normalize, Norm Ref Lvl
Normalized Reference Level Position
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRPosition <integer>
:DISPlay:WINDow:TRACe:Y[:SCALe]:NRPosition?
Selects the position of the normalized reference level. The top and bottom graticule
lines correspond to 10 and 0, respectively.
NOTE
See command
:CALCulate:NTData[STATe] OFF|ON|0|1
Factory Preset
and *RST:
10
Range:
integer
Front Panel
Access:
View/Trace, Normalize, Norm Ref Posn
Содержание E7401A
Страница 36: ...36 Commands Alphabetical Listing ...
Страница 92: ...92 Chapter 2 Status Registers Use Status Registers to Determine the State of Analyzer Events and Conditions ...
Страница 185: ...185 4 Programming Command Cross References ...
Страница 283: ...Chapter 5 283 Language Reference SENSe DETector Subsection Figure 5 4 Auto Rules of Detector Selection ...
Страница 339: ...339 6 Agilent 8590 EMC Analyzers Programming Conversion Guide ...