background image

 4

Chapter 1

 

General Information

Option 028 Hard Disk Drive Kit

Option 028 Hard Disk Drive Kit

The purpose of this kit is to install the option 028 into the Agilent E5071C. 

Contents of the Agilent E5071CU Option 028 Upgrade Kit

The kit includes the parts listed in Table 1-1. If anything is missing, contact the nearest 
Agilent Technologies office.

Table 1-1 

Contents of the Agilent E5071CU Option 028 Upgrade 
Kit

Description

Qty.

Upgrade Manual (This manual)

1

Hard Disk for E5071C

1

Содержание E5071CU 028

Страница 1: ...Agilent E5071CU Option 028 Hard Disk Drive Kit User s Guide First Edition Manufacturing No E5071 90133 August 2011 ...

Страница 2: ...rated at reprint do not cause the date to change The manual part number changes when extensive technical changes are incorporated August 2011 First Edition Typeface Conventions Sample bold Boldface type is used when a term is defined or emphasis Sample Italic Italic type is used for emphasis key Indicates a hardkey key on the front panel or external keyboard labeled Sample key may be omitted Sampl...

Страница 3: ...3 1 General Information This chapter describes the Hard Disk Drive Kit contents of the kit and required tools to install the Option 028 into the Agilent E5071C ...

Страница 4: ...tion 028 into the Agilent E5071C Contents of the Agilent E5071CU Option 028 Upgrade Kit The kit includes the parts listed in Table 1 1 If anything is missing contact the nearest Agilent Technologies office Table 1 1 Contents of the Agilent E5071CU Option 028 Upgrade Kit Description Qty Upgrade Manual This manual 1 Hard Disk for E5071C 1 ...

Страница 5: ...5 2 Option Installation This chapter describes the procedure to install the Option 028 into the Agilent E5071C ...

Страница 6: ...ch Screen Update Procedure for Hard Disk This procedure is for screw type hard disk Installation Requirement 5188 4426 Key for Hard Disk Carrier Backing Up Option Data Step 1 Copy RECOVERY E LICENSE gen lic from the original HDD to USB pen drive NOTE The license keys are kept in a text file lic located at RECOVERY E LICENSE gen lic When you change the removal hard disk copy the file as back up Eve...

Страница 7: ...Step 4 Press the button under the screw Step 5 Remove the hard disk Replacement Procedure Step 1 Insert the hard disk into the slot Step 2 Turn the screw in anti clockwise direction until it is locked Step 3 Turn the key to the right until it is locked Step 4 Remove the key Step 5 Turn on the E5071C Restoring Option Data Step 1 Paste the gen lic file to RECOVERY E LICENSE on Spare HDD from the USB...

Страница 8: ...p 2 Press Step 3 Press Load Project on the softkey menu Step 4 A dialog box appears to allow you to select a program to be loaded Select RestoreSysCorFile vba from the D Agilent Service folder and then press Open button Step 5 Press Step 6 RestoreSysCorFile dialog box appears Figure 2 1 Click OK Figure 2 1 RestoreSysCorFile Dialog Box CAUTION Once the program has ended restart the E5071C to enable...

Страница 9: ...lick My Computer Step 2 Select Properties The System Properties dialog appears Step 3 Select Hardware tab Step 4 Click Device Manager Step 5 Select Ports COM LPT Step 6 Check if two Communication Port COM1 are displayed Step 7 Right click and select Uninstall Figure 2 2 for both the COM1 ports Figure 2 2 Device Manager Step 8 The Confirm Device Removal message appears Click OK Step 9 Reboot the in...

Страница 10: ...n The touch screen calibration screen Figure 2 3 appears Figure 2 3 Touch Panel Calibration Screen Step 5 Touch the x mark on the upper left with your finger The mark x appears also on the lower left upper right and lower right Touch the x marks in that order with your finger Touching the four locations described above with your finger automatically concludes the touch screen calibration NOTE With...

Страница 11: ......

Страница 12: ... Agilent Technologies Inc 2011 Printed in Malaysia First edition August 2011 E5071 90133 E5071 90133 www agilent com Agilent Technologies ...

Отзывы: