5. TRACE NOISE TEST
5. TRACE NOISE TEST
Description
This test checks the trace noise level for each test port of E5071C. The trace noise level is
quantified by performing a “through” measurement 32 times from 9 kHz to 8.5 GHz range,
with a cable connected between two ports. The trace noise level for S11, S22, S33 and S44
are quantified by performing a “short” measurement 32 times from 9 kHz to 8.5 GHz
range, with a cable (terminated with “Short” device) connected to each ports(1,2,3, and 4).
Standard deviation of the measured values at each frequency is calculated and, then
translated into a noise level expressed in dB rms / deg rms.
Test equipment
20 inch 50 ohm cable
Agilent p/n 5062-6691
N-SMA Adapter
Agilent p/n 1250-2879
Type N, 50 ohm Short termination (1ea)
Agilent p/n 85032-60009
Содержание E5071C
Страница 14: ...14 Contents ...
Страница 168: ...168 Chapter2 Performance Test E5071C Performance Test Record ...
Страница 212: ...212 Chapter4 Troubleshooting To configure the CPU Mother Board Assembly and BIOS ...
Страница 262: ...262 Chapter5 Replaceable Parts Power Cables and Plug Configurations ...
Страница 326: ...326 Chapter6 Replacement Procedure Probe Power and Power Switch Board Replacement ...
Страница 332: ...332 Chapter7 Post Repair Procedures Post Repair Procedures ...
Страница 360: ...360 AppendixB System Recovery Calibration of the Touch Screen ...
Страница 364: ...364 AppendixC Firmware Update Firmware Update ...
Страница 365: ...365 D Power Requirement D Power Requirement ...
Страница 367: ...Appendix D 367 Power Requirement Preparation for Power Supply D Power Requirement Figure D 1 Power cable options ...