22
Chapter 1
General Information
Power Meter Accuracy Test
Power Meter Accuracy Test
This test is intended for power meters used in testing the E5071C. The “Dynamic Accuracy
Test” requires the use of a power meter that has been calibrated to a higher accuracy than
the standard power meter.
Power meters with options G12 and H12 specify an improved instrument accuracy over a
limited power range. (These power meters do not contain unique hardware.) A power
meter may be returned to the factory to have one of these options added to an existing
power meter or to renew the calibration for one of these options.
This test procedure is an alternative to returning the power meter to the factory. When a
power meter passes this test, it is considered to be
calibrated
for the G12 or H12 option
even though it has not been returned to the factory.
Power Meters That Can Be Tested Using This Procedure
This procedure assume that the E4419B power meter is being tested.
NOTE
It is recommended that the revision number for the power meter “Main Firmware” be
Ax.03.00 or higher.
Equipment Used for the Power Meter Accuracy Test
Equipment Type
Recommended Model
or Part Number
Alternate
Model or Part Number
Range calibrator
Agilent 11683A
None
Precision digital
voltmeter
Agilent 3458A
Any with the required
accuracy and resolution
*1.Required accuracy and resolution at the following voltage levels:
14 mV input: 0.0100% accuracy, 10 nV resolution
0.14 V input: 0.0050% accuracy, 100 nV resolution
0.45 V input: 0.0030% accuracy, 100 nV resolution
*1
Power sensor
cable
Agilent 11730A, p/n
8120-8319
Any equivalent
Cable, BNC, 50
Ω
Any
Any
Adapter for
connecting BNC
cable to DVM
inputs
Any
Any
Description of the Test
The power meter accuracy is verified for various power inputs and the actual readings are
recorded in a test record. A range calibrator is used to provide the reference inputs.
Содержание E5071C
Страница 14: ...14 Contents ...
Страница 168: ...168 Chapter2 Performance Test E5071C Performance Test Record ...
Страница 212: ...212 Chapter4 Troubleshooting To configure the CPU Mother Board Assembly and BIOS ...
Страница 262: ...262 Chapter5 Replaceable Parts Power Cables and Plug Configurations ...
Страница 326: ...326 Chapter6 Replacement Procedure Probe Power and Power Switch Board Replacement ...
Страница 332: ...332 Chapter7 Post Repair Procedures Post Repair Procedures ...
Страница 360: ...360 AppendixB System Recovery Calibration of the Touch Screen ...
Страница 364: ...364 AppendixC Firmware Update Firmware Update ...
Страница 365: ...365 D Power Requirement D Power Requirement ...
Страница 367: ...Appendix D 367 Power Requirement Preparation for Power Supply D Power Requirement Figure D 1 Power cable options ...