522
Chapter 15
Measurement Examples
Evaluating Transmission Characteristics of a Front End Module
1. Determining Measurement Conditions
In this example, perform measurement under the measurement conditions in Table 15-9.
Table 15-10
Sweep conditions
Channel
Start
frequency
Stop
frequency
NOP
Test port assignment
Control line
Meas.
parameter
Calibration
Type
Port
1
400 MHz
1.4 GHz
201
Port 1 - Port A
Port 2 - Port T1
Port 3 - Port R1+
Port 4 - Port R1-
Line 1: Low
Line 2: High
S12
Full 2-Port
1,2
2
880 MHz
1 GHz
101
Line 1: Low
Line 2: Low
Sds21
Full 3-Port
1,3,4
3
1.34 GHz
2.34 GHz
201
Port 1 - Port A
Port 2 - Port T2
Port 3 - Port R2+
Port 4 - Port R1-
Line 1: High
Line 2: Low
S12
Full 2-Port
1,2
4
1.665 GHz 2.015 GHz
101
Line 1: Low
Line 2: Low
S31
Full 2-Port
1,3
2. Setting Channel Window Allocation
Set the screen to split into 2 rows and 2 columns to assign channel windows after preset.
Setting Description
Key Operation
Execute preset
-
OK
Allocate channel windows
-
Allocate Channels
-
3. Setting the Test Ports
Step 1.
Display the E5091A setup menu and select the 9-port model for ID1.
Setting Description
Key Operation
Display the E5091A setup menu
-
Multiport Test Set Setup
Select the 9-port model for ID1
Test Set 1 - Select Test Set - E5091_9
Step 2.
Display the E5091A properties.
Setting Description
Key Operation
Display the E5091A properties
Property
“10. Executing Measurement” on page 527
Execute the measurement and perform auto scale.
Table 15-9
Evaluation Procedure for 6-port Front End Module
Procedure
Description
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...