![Agilent Technologies E5070B Скачать руководство пользователя страница 387](http://html.mh-extra.com/html/agilent-technologies/e5070b/e5070b_user-manual_2868065387.webp)
Chapter 11
381
Limit Test
Limit Test
11
. Lim
it T
es
t
Limit line offset
By adding a certain offset to the limit value, you can adjust the limit line so that it conforms
to the device output.
Step 1.
Press
or
to activate the channel to which you want to apply the
limit test function.
Step 2.
Press
or
to activate the trace to which you want to apply the limit test
function.
Step 3.
Press
to display the Analysis menu.
Step 4.
Press
Limit Test
to display the softkeys for the limit test.
Step 5.
Press
Limit Line Offsets
to display the limit line offset function menu. The following
functions correspond to each softkey.
Softkey
Function
Stimulus Offset
Adds a certain offset to the stimulus value of the entire segment in the
limit table. (Stimulus offset)
Amplitude Offset
Adds a certain offset to the response value of the entire segment in the
limit table. (Amplitude offset)
Marker Amp. Ofs.
Adds the amplitude offset by the same amount as the retrieved value of the
active marker. You can confirm the current value set for the amplitude
offset by pressing
Amplitude Offset
. (Marker amplitude offset)
Return
Ends the limit line offset menu.
Figure 11-5
Stimulus offset
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...