Chapter 5
247
Making Measurements
Setting the low-latency external trigger mode
5. M
a
king
Me
asur
eme
n
t
Step 1.
Press
.
Step 2.
Press
Ext Trig Delay
.
Step 3.
Enter an external trigger delay time.
External trigger delay time and point trigger interval
External trigger pulses, supplied until the next measurement becomes ready after the start
of a one-point measurement, are ignored, and the next trigger is generated by a pulse
supplied after the completion of the one-point measurement.
The time until the next trigger can be accepted after the start of a one-point measurement
depends on the IFBW and other settings of the analyzer. For example, in the case of a
frequency’s zero-span measurement, the time until the next measurement is ready after the
start of a one-point measurement is obtained by dividing the time required for a single
sweep in On Sweep mode, instead of On Point mode, by the number of measurement
points. If you use the point trigger function with external trigger pulses that are wider than
this time, point trigger measurement is performed at each pulse input.
The figure below shows the timing chart of an external trigger when the point trigger
function is on.
Figure 5-1
Timing chart of external trigger (trigger event = on, trigger source = external,
low-latency external trigger mode = on)
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...