Chapter 4
179
Calibration
Confidence Check on Calibration Coefficients Using ECal
4. Cal
ibra
tion
Confidence Check on Calibration Coefficients Using ECal
NOTE
This function is available with firmware version 3.50 or higher.
Using the ECal module, the E5070B/E5071B lets you verify the obtained calibration
coefficients to determine whether correct measurement is possible with them.
The E5070B/E5071B can set ECal to the state used to verify the measurement parameters
and then copy the appropriate characteristics of that verification state to the memory trace
from the ECal's built-in memory. This is done according to the measurement parameters of
the active trace of the active channel. While measuring ECal in this specified state, you can
compare the measurement results with those of the E5070B/E5071B and with the
appropriate measurement results stored in ECal in several different ways. These include
simultaneously displaying the data and memory traces or displaying the math operation
results between the data and memory traces. This enables you to verify the correctness of
measurement for each measurement parameter when the obtained calibration coefficients
are used.
Operational procedure
Follow these steps to verify the reliability of the calibration coefficients.
Step 1.
Connect the USB cable between the USB port of the ECal module and that of the
E5070B/E5071B. You can make this connection while the E5070B/E5071B’s power is
ON.
Step 2.
Press
or
to select the channel for which you want to perform the
verification.
Step 3.
Press
.
Step 4.
Select the S-parameter you want to verify.
NOTE
You cannot verify the mixed mode S-parameter.
Step 5.
Connect the test ports of the E5070B/E5071B corresponding to the selected S-parameter
(for example, ports 1 and 2 when the S-parameter is S21) and the ports of the ECal module.
NOTE
If you don’t use all of the ECal module’s ports, connect terminations to the unused
ports.
NOTE
You can use any of the Ecal’s ports for this connection. The connected port is automatically
detected and measurement is performed.
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...