122
Chapter 4
Calibration
Full 3-Port Calibration
Figure 4-18
Connecting standards in full 3-port calibration
Procedure
Step 1.
Press
or
to select the channel for which you want to perform the
calibration.
Step 2.
Press
.
Step 3.
Press
Calibrate
.
Step 4.
Press
3-Port Cal
.
Step 5.
Press
Select Ports
.
Step 6.
Select the test ports on which you will perform full 3-port calibration. (In the procedure
below, the selected test ports are denoted as x, y, and z.)
Softkey
Function
1-2-3
Selects test ports 1, 2, and 3
1-2-4
Selects test ports 1, 2, and 4
1-3-4
Selects test ports 1, 3, and 4
2-3-4
Selects test ports 2, 3, and 4
NOTE
An asterisk (*) in the upper-right of the softkey indicates that the existing calibration
coefficient will be cleared if you select the test port and execute acquisition of the
Содержание E5070B
Страница 6: ......
Страница 30: ...24 Contents ...
Страница 34: ...28 Chapter1 Precautions Before contacting us ...
Страница 286: ...280 Chapter6 Data Analysis Using the Equation Editor ...
Страница 430: ...424 Chapter12 Optimizing Measurements Performing a Segment by Segment Sweep segment sweep ...
Страница 479: ...473 14 Controlling E5091A 14 Controlling E5091A This chapter explains how to control the E5091A multiport test set ...
Страница 538: ...532 Chapter15 Measurement Examples Executing Power Calibration ...
Страница 604: ...598 AppendixA Manual Changes Manual Changes Figure A 12 Two port Touchstone file Figure A 13 Three port Touchstone file ...
Страница 634: ...628 AppendixB Troubleshooting Warning Message ...
Страница 732: ...726 AppendixD Softkey Functions Trigger Menu ...
Страница 740: ...734 AppendixE General Principles of Operation Data Processing ...
Страница 760: ...754 AppendixF Replacing the 8753ES with the E5070B E5071B Comparing Functions ...