296
7
SCPI Command Reference
:SENSe:ATTenuation:LEVel
Parameter
Equivalent key
No equivalent key is available on the front panel.
:SENSe:ATTenuation:LEVel
Syntax
:SENSe:ATTenuation:LEVel <numeric>
:SENSe:ATTenuation:LEVel?
Description
Sets/Reads the input attenuator level
Parameter
When the E5052A is used stand-alone, or with the downconverter turned off, or with
the downconverter on and the RF input is set to 'E5052A Direct'
When the downconverter is turne on and the RF input is set to 'Downconverter'
Equivalent key
PN Menu -> Attenuator -> Input Attenuator
SP Menu -> Attenuator -> Input Attenuator
FP Menu -> Attenuator -> Input Attenuator
TR Menu -> Attenuator -> Input Attenuator
USER Menu -> Attenuator -> Input Attenuator
<String>
Range
-
Preset value
-
Unit
-
Resolution
-
<Description>
Range
0 to 35
Preset value
5
Unit
dB
Resolution
5
<Description>
Range
10
Preset value
10
Unit
dB
Resolution
-
Содержание E5052A
Страница 4: ...4 ...
Страница 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Страница 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Страница 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Страница 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Страница 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...
Страница 476: ...476 AppendixB Status Reporting System Status Register Structure Figure B 3 Status register structure 1 of 6 ...
Страница 478: ...478 AppendixB Status Reporting System Status Register Structure Figure B 5 Status register structure 3 of 6 ...
Страница 480: ...480 AppendixB Status Reporting System Status Register Structure Figure B 7 Status register structure 5 of 6 ...