236
7
SCPI Command Reference
:DISPlay:FP[1-1]:Y[:SCALe]:DIVisions
Parameter
Equivalent key
FP Menu -> Scale -> Reference Position
:DISPlay:FP[1-1]:Y[:SCALe]:DIVisions
Syntax
:DISPlay:FP[1-1]:Y[:SCALe]:DIVisions <numeric>
:DISPlay:FP[1-1]:Y[:SCALe]:DIVisions?
Description
# of Y division
Parameter
Equivalent key
FP Menu -> Scale -> Divisions
:DISPlay:MAXimize
Syntax
:DISPlay:MAXimize {ON|OFF|1|0}
:DISPlay:MAXimize?
Description
Maximizes active instrument window
Parameter
Equivalent key
No equivalent key is available on the front panel.
<Numeric>
Range
0 to 30
Preset value
5
Unit
Div
Resolution
-
<Numeric>
Range
4 to 30
Preset value
10
Unit
-
Resolution
2
Description
ON or 1(Preset
value)
Maximize active instrument window mode to 'ON'
OFF or 0
Maximize active instrument window mode to 'OFF'
Содержание E5052A
Страница 4: ...4 ...
Страница 28: ...28 1 Making Effective Use of This Manual How to Use This Manual ...
Страница 51: ...3 51 Making a Measurement Trigger System 3 Making a Measurement Figure 3 1 Trigger system ...
Страница 58: ...58 3 Making a Measurement Waiting for End of measurement ...
Страница 76: ...76 4 Reading Writing Measurement Data Entering Data in a Trace ...
Страница 82: ...82 Chapter5 Saving and Recalling File Management Managing Files ...
Страница 476: ...476 AppendixB Status Reporting System Status Register Structure Figure B 3 Status register structure 1 of 6 ...
Страница 478: ...478 AppendixB Status Reporting System Status Register Structure Figure B 5 Status register structure 3 of 6 ...
Страница 480: ...480 AppendixB Status Reporting System Status Register Structure Figure B 7 Status register structure 5 of 6 ...