52
Chapter 3
Basic Operation
Configuring the RF Output
7. Press
Freq Offset
>
1
>
MHz
.
This enters a 1 MHz offset. The
FREQUENCY
area displays
2.000 000 00 MHz
, which is the
frequency output by the hardware (701 MHz) minus the reference frequency (700 MHz) plus the offset
(1 MHz). The
OFFS
indicator is activated. The frequency at the RF OUTPUT connector is still
701 MHz.
Setting the RF Output Amplitude
1. Press
Preset
.
2. Observe the
AMPLITUDE
area of the display.
The display reads the minimum power level of the signal generator. This is the normal preset RF output
amplitude.
3. Press
RF On/Off
.
The display annunciator changes from
RF OFF
to
RF ON
. The RF signal is now being output at the
minimum power level at the RF OUTPUT connector.
4. Press
Amplitude
>
–20
>
dBm
.
This changes the amplitude to
−
20 dBm. The new
−
20 dBm RF output power is now displayed in the
AMPLITUDE
area of the display and also in the active entry area.
Amplitude is still the active function until you press another front panel function key. You can also
change the amplitude using the up and down arrow keys and the knob.
Setting the Amplitude Reference and Amplitude Offset
The following procedure sets the RF output power as an amplitude reference to which all other amplitude
parameters are relative. The amplitude initially shown on the display will be 0 dB (the power output by the
hardware minus the reference power). Although the display changes, the output power does not change. Any
subsequent power changes are shown as incremental or decremental to 0 dB.
1. Press
Preset
.
2. Press
Amplitude
>
–20
>
dBm
.
3. Press
More (1 of 2)
>
Ampl Ref Set
.
This activates the amplitude reference mode and sets the current output power (
−
20 dBm) as the
reference value. The
AMPLITUDE
area displays
0.00 dB
, which is the power output by the hardware
(
20 dBm) minus the reference value (
−
20 dBm). The
REF
indicator is activated and the
Ampl Ref Off On
softkey has toggled to On.
Содержание E4428C
Страница 22: ...Contents xxii ...
Страница 107: ...Chapter 3 83 Basic Operation Using Security Functions Figure 3 6 ESG Screen with Secure Display Activated ...
Страница 182: ...158 Chapter 4 Basic Digital Operation Using Waveform Clipping Figure 4 22 Rectangular Clipping ...
Страница 183: ...Chapter 4 159 Basic Digital Operation Using Waveform Clipping Figure 4 23 Reduction of Peak to Average Power ...
Страница 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Страница 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Страница 229: ...205 6 Analog Modulation ...
Страница 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Страница 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Страница 287: ...263 9 BERT This feature is available only in E4438C ESG Vector Signal Generators with Option 001 601or 002 602 ...
Страница 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Страница 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Страница 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Страница 454: ...430 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Concepts Figure 15 9 Uplink Channel Structure ...
Страница 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Страница 667: ...643 18 Troubleshooting ...
Страница 700: ...Index 676 Index ...