82
Chapter 3
Basic Operation
Using Security Functions
The
Enter Secure Mode
softkey changes to
Secure Mode Activated
.
If Your Instrument is Not Functioning
If the instrument is not functioning and you are unable to use the security functions, you must physically
remove the processor board and optional hard disk, if installed, from the instrument. Once these assemblies
are removed, proceed as follows:
For the processor board, choose one of the following options:
•
Discard the processor board and send the instrument to a repair facility. A new processor board will be
installed and the instrument will be repaired and calibrated. If the instrument is still under warranty, you
will not be charged for the new processor board.
•
If you have another working instrument, install the processor board into that instrument and erase the
memory. Then reinstall the processor board back into the non-working instrument and send it to a repair
facility for repair and calibration. If you discover that the processor board does not function in the
working instrument, discard the processor board and note that it caused the instrument failure on the
repair order. If the instrument is still under warranty, you will not be charged for the new processor
board.
For the optional hard disk, choose one of the following options:
•
Discard the hard disk and send the instrument to a repair facility. Indicate on the repair order that the
hard disk was removed and must be replaced. A new hard disk will be installed and the instrument will
be repaired and calibrated. If the instrument is still under warranty, you will not be charged for the new
hard disk.
•
Keep the hard disk and send the instrument to a repair facility. When the instrument is returned, reinstall
the hard disk.
For procedures on removing and replacing the processor board and hard disk, refer to the
E4428C/38C ESG
Signal Generators Service Guide
.
Using the Secure Display
This function prevents unauthorized personnel from reading the instrument display and tampering with the
current configuration through the front panel. The display is blanked , except for the message
*** SECURE
DISPLAY ACTIVATED ***
(see
), and the front panel keys are disabled. Once this
function is activated, the power must be cycled to re-enable the display and front panel keys.
To apply this function, press
Utility
>
Display
>
More (1 of 2)
>
Activate Secure Display
>
Confirm Secure Display
Содержание E4428C
Страница 22: ...Contents xxii ...
Страница 107: ...Chapter 3 83 Basic Operation Using Security Functions Figure 3 6 ESG Screen with Secure Display Activated ...
Страница 182: ...158 Chapter 4 Basic Digital Operation Using Waveform Clipping Figure 4 22 Rectangular Clipping ...
Страница 183: ...Chapter 4 159 Basic Digital Operation Using Waveform Clipping Figure 4 23 Reduction of Peak to Average Power ...
Страница 224: ...200 Chapter 4 Basic Digital Operation Creating and Using Bit Files ...
Страница 228: ...204 Chapter 5 AWGN Waveform Generator Configuring the AWGN Generator ...
Страница 229: ...205 6 Analog Modulation ...
Страница 276: ...252 Chapter 7 Digital Signal Interface Module Operating the N5102A Module in Input Mode ...
Страница 286: ...262 Chapter 8 Bluetooth Signals Turning On a Bluetooth Signal ...
Страница 287: ...263 9 BERT This feature is available only in E4438C ESG Vector Signal Generators with Option 001 601or 002 602 ...
Страница 330: ...306 Chapter 9 BERT Verifying BERT Operation ...
Страница 366: ...342 Chapter 10 CDMA Digital Modulation IS 95A Modulation ...
Страница 394: ...370 Chapter 12 Multitone Waveform Generator Applying Changes to an Active Multitone Signal ...
Страница 454: ...430 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Concepts Figure 15 9 Uplink Channel Structure ...
Страница 468: ...444 Chapter 15 W CDMA Digital Modulation for Component Test W CDMA Frame Structures ...
Страница 667: ...643 18 Troubleshooting ...
Страница 700: ...Index 676 Index ...