Chapter 3
53
Operation Verification
Test Tables
−
65
−
66.5
−
63.5
−
75
−
76.5
−
73.5
−
85
−
86.5
−
83.5
Table 3-10
FM Accuracy
Frequency
(MHz)
Deviation (kHz)
Limits (kHz)
Lower
Measured
Upper
500.001
100 kHz
96.48
103.52
750
100 kHz
96.48
103.52
1000
100 kHz
96.48
103.52
Table 3-9
Power Level Accuracy Setup 2
Frequency
Setting
Power
Level
Setting
(dBm)
Power
Meter
Reading
for
−
45 dBm
Spectrum
Analyzer
Marker
(dB)
Lower
Limit
(dBm)
Actual
Power Level
(dBm)
Upper
Limit
(dBm)