Operation Verification
Power Level Accuracy Check
Power Level Accuracy Check
Performing this check will provide a high level of confidence that the signal generator’s power level
circuitry is functioning correctly. This check does not test the signal generator to warranted specifications.
Test points have been reduced and the limits are degraded in order to minimize measurement time and take
into account a broad range of measurement uncertainties.
In order to have the signal generator tested to warranted specifications, a complete power level accuracy
performance test is required. If the complete performance test is needed and you are unable to perform it,
contact your nearest Agilent Technologies service center for information concerning calibration. Refer to
“Contacting Agilent Sales and Service Offices” on page 21
Connect the Test Equipment for Setup 1
Figure 3-2
Power Level Accuracy Equipment Setup 1
Configure the Signal Generator
1. Press
Preset
.
2. Press
Frequency
. Enter 277 using the numeric keypad and press the
kHz
terminator softkey.
3. Press
Amplitude
. Enter 13 using the numeric keypad and press the
dBm
terminator softkey.
4. Press
Mod On/Off
. The
MOD OFF
annunciator is displayed.
5. Press
RF On/Off
. The
RF ON
annunciator is displayed.
Содержание E4428C ESG
Страница 7: ...1 1 Safety Information ...
Страница 12: ...6 Chapter 1 Safety Information General Safety Considerations ...
Страница 13: ...7 2 Getting Started ...
Страница 28: ...22 Chapter 2 Getting Started Contacting Agilent Sales and Service Offices ...
Страница 29: ...23 3 Operation Verification ...
Страница 61: ...55 4 Regulatory Information ...
Страница 64: ...58 Chapter 4 Regulatory Information Regulatory Information misc ...