4
Error Messages
156
E4419B Power Meter User’s Guide
Error Messages
–101
Invalid character
An invalid character was found in the command string. You may have inserted a character such
as #, $, or % in the command header or within a parameter.
For example,
LIM:LOW O#.
–102
Syntax error
Invalid syntax was found in the command string.
For example,
LIM:CLE:AUTO, 1
or
LIM:CLE: AUTO 1
.
–103
Invalid separator
An invalid separator was found in the command string. You may have used a comma instead of
a colon, semicolon, or blank space; or you may have used a blank space instead of a comma.
For example,
OUTP:ROSC,1
.
–105
GET not allowed
A Group Execute Trigger (
GET
) is not allowed within a command string.
–108
Parameter not allowed
More parameters were received than expected for the command. You may have entered an
extra parameter, or added a parameter to a command that does not accept a parameter.
For example,
CAL 10
.
–109
Missing parameter
Fewer parameters were received than expected for the command. You omitted one or more
parameters that are required for this command. For example,
AVER:COUN
.
–112
Program mnemonic too long
A command header was received which contained more than the maximum 12 characters
allowed.
For example,
SENSeAVERageCOUNt 8
.
–113
Undefined header
A command was received that is not valid for this power meter. You may have misspelled the
command, it may not be a valid command or you may have the wrong interface selected. If you
are using the short form of the command, remember that it may contain up to four letters.
For example,
TRIG:SOUR IMM
.
–121
Invalid character in number
An invalid character was found in the number specified for a parameter value.
For example,
SENS:AVER:COUN 128#H
.
Содержание E4419B
Страница 1: ...Agilent Technologies Agilent E4419B Power Meter User s Guide ...
Страница 14: ...xiv E4419B Power Meter User s Guide ...
Страница 22: ...xxii E4419B Power Meter User s Guide ...
Страница 44: ...2 Power Meter Operation 22 E4419B Power Meter User s Guide Contacting Agilent Technologies 109 ...
Страница 117: ...Power Meter Operation 2 E4419B Power Meter User s Guide 95 Figure 2 22 Save Recall Screen ...
Страница 136: ...2 Power Meter Operation 114 E4419B Power Meter User s Guide ...
Страница 174: ...3 Menu Reference 152 E4419B Power Meter User s Guide ...
Страница 175: ...153 User s Guide Agilent Technologies 4 Error Messages Introduction on page 154 Error Messages on page 156 ...
Страница 184: ...4 Error Messages 162 E4419B Power Meter User s Guide ...