Chapter 6
385
Error Messages
Error Message Descriptions
Error Message Descriptions
Messages with No Numbers
Unnumbered messages are for operator information only and do not
appear in any error queue.
Description
Acquiring Data...
A warning used when the data acquisition time is long
enough to be noticeable.
AFUN not implemented
Awaiting Trigger, no AUTO Trig
Auto Trig is off and a trigger has not been detected for
more than 4 seconds.
Break freq > FFT filter edge - clipping to %f kHz
Correction off
Data Acquisition FIFO_OVERFLOW, use AUTO
DataPacking...”
Data acquisition malfunction; need to use auto data
packing to resolve.
GSM Hopping enabled, waiting for valid burst
When GSM hopping is enabled, this indicates that a
valid GSM burst has not yet been found.
IF synthesizer unlocked
LAN external loopback test failed
This message will appear during boot-up if the
instrument is not connected to a LAN cable. You can
ignore the message if you are not using the LAN.