background image

Installation Note

Part Number E4400-90550
Printed in USA May 2002
Supersedes April 2002

Agilent Technologies ESG Vector Signal Generator 
Upgrade Kit for Option 002 
(Internal Baseband Generator with 32 Msample Memory) 
Kit Part Number E4400-60598

Содержание E4400-60598

Страница 1: ...rt Number E4400 90550 Printed in USA May 2002 Supersedes April 2002 Agilent Technologies ESG Vector Signal Generator Upgrade Kit for Option 002 Internal Baseband Generator with 32 Msample Memory Kit Part Number E4400 60598 ...

Страница 2: ...d to this material including but not limited to the implied warranties of merchantability and fitness for a particular purpose Agilent Technologies shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material Copyright 2002 Agilent Technologies Inc ...

Страница 3: ...ion 1EM rear panel connector setup Installation includes the following major steps 1 Check Functionality 2 Remove the instrument covers 3 Install the A7 Internal Baseband Generator board 4 Install new firmware 5 Activate Option 002 6 Re assemble the signal generator 7 Calibrate the signal generator Product Affected E4438C Serial Numbers All Options To Be Performed By X Agilent Technologies Service...

Страница 4: ...an damage or destroy electronic components All work on electronic assemblies should be performed at a static safe workstation Item Quantity Description Part Number 1 1 Installation Note E4400 90550 2 1 ESG C Firmware Upgrade Kit E4400 60597 3 1 Option 002 Label 7121 1232 4 1 A7 Internal Baseband Generator 32 Msample Memory E4400 60501 5 1 W10 Ribbon Cable 8121 0679 For Non Option 1EM signal genera...

Страница 5: ...t five minutes 2 Run the instrument self test by pressing Utility Instrument Info Help Mode Self Test Run Complete Self Test Upon completion a summary of the self test will be displayed Use the service guide to troubleshoot any failures detected by the test NOTE Some circuits may require up to 50 minutes to warm up before passing the self test If self tests continue to fail after 50 minutes of war...

Страница 6: ...o strap handles 1 by loosening the screws 3 Using a T 15 driver remove the center screws 3 on the four rear panel feet 2 4 Remove the four bottom feet 5 and 6 from the cover by pushing and pulling the tab 5 Slide the outer cover 4 off the frame Removing the Inner Top Cover Refer to Figure 1 1 Using a T 10 driver remove the ten screws 8 from the inner top cover 7 2 Remove the inner top cover ...

Страница 7: ...7 Figure 1 Outer and Inner Cover Removal ...

Страница 8: ...8 Figure 2 ...

Страница 9: ...Generator Board Refer to Figure 2 for this procedure 1 Locate Top SLOT 1 2 Insert the A7 assembly into the front and rear card cage guides 3 Install the A7 Assembly onto the A23 motherboard 4 Install the W78 cable to the rear panel slot labeled BBG REF IN Use a lock washer and a hex nut to secure one end of the cable to the rear panel Torque to 9 in lb 5 Connect the loose end of the W78 cable to A...

Страница 10: ... following A7 Baseband Generator adjustments in the order listed Self Test I Q Gain Offset Quadrature Calibration I Q Impairment Calibration Verifying Instrument Calibration To ensure proper calibration of the A7 Baseband Generator use the following performance tests in the order listed Self Test Harmonic Subharmonic and Nonharmonic Spurious Signals Digital Modulation Level Accuracy Internal Digit...