Chapter 4 Remote Interface Reference
Output Setting and Operation Commands
86
CURRent:STEP? [DEFault]
Return the value of the step size currently specified. The returned parameter is a
numeric value. ‘‘
DEFault
’’ gives the minimum resolution of the step size in unit of
amps.
CURRent:TRIGgered {
<current>
| MINimum | MAXimum}
Program the pending triggered current level. The pending triggered current level is a
stored value that is transferred to the output terminals when a trigger occurs. A pending
triggered level is not affected by subsequent
CURRent
commands.
CURRent:TRIGgered? [MINimum | MAXimum]
Query the triggered current level presently programmed. If no triggered level is
programmed, the
CURRent
level is returned.
Example
The following program segments show how to use the
CURR
UP
or
CURR
DOWN
command to increase or decrease the output current with the
CURR
:
STEP
command.
‘‘
CURR
:
STEP
0.01
’’
Set the step size to 0.01 A
‘‘
CURR
UP
’’
Increase the output current
‘‘
CURR
:
STEP
0.02
’’
Set the step size to 0.02 A
‘‘
CURR
DOWN
’’
Decrease the output current
INSTrument[:SELect] {OUTPut1 | OUTPut2 | OUT1 | OUT2}
Select the output to be programmed one of the two outputs by the output identifier.
The outputs of the power supply are considered as two logical instruments. The
INSTrument
command provides a mechanism to identify and select an output.
When one output is selected, the other output is unavailable for programming until
selected. The commands which are affected by the
INSTrument
command are
output setting command (
SOURce
), measurement command (
MEASure
), and
calibration command (
CALibration
).
INSTrument[:SELect]?
Return the currently selected output by the
INSTrument{:SELect]
or
INSTrument:NSELect
command. The returned value is ‘‘OUTP1’’ or ‘‘OUTP2’’.
INSTrument:NSELect {1 | 2}
Select the output to be programmed one of the two outputs by
a numeric value
instead
of the output identifier used in the
INSTrument:NSELect
or
INSTrument[:SELect]
command. ‘‘1’’ selects output1, and ‘‘2’’ selects
output2.
Содержание E3646A
Страница 2: ...2...
Страница 16: ...16 Contents Contents...
Страница 17: ...1 Quick Start...
Страница 27: ...2 General Information...
Страница 39: ...3 Front Panel Operation and Features...
Страница 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Страница 73: ...4 Remote Interface Reference...
Страница 125: ...5 Error Messages...
Страница 136: ...Chapter 5 Error Messages Calibration Errors 136...
Страница 137: ...6 Application Programs...
Страница 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Страница 149: ...7 Tutorial...
Страница 160: ...Chapter 7 Tutorial Remote Programming 160...
Страница 161: ...8 Specifications...
Страница 168: ...Chapter 8 Specifications Supplemental Characteristics 168 Figure 8 2 Dimensions for Rack mounting...
Страница 169: ...Appendix Service Information...
Страница 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Страница 210: ......