Appendix Service Information
Self-Test Procedures
176
Self-Test Procedures
Power-On Self-Test
Each time the power supply is powered on, a set of self-tests are performed. These
tests check that the minimum set of logic and measurement hardware are functioning
properly. Failures during the power-on self-test utilize error codes 601 through 604
and 624 through 632.
Complete Self-Test
Hold any front panel key except the ‘‘View’’ key for more than 5 seconds while turning
on the power to perform a complete self-test. The power supply beeps when the test
starts. The tests are performed in the order shown below.
601
Front Panel Does not respond
The main controller U17 attempts to establish serial
communications with the front panel controller U1 on the front panel board. During
this test, the U1 turns on all display segments. Communication must function in both
directions for this test to pass. If this error is detected during power-on self-test, the
power supply will beep twice. This error is only readable from the remote interface.
602
RAM read/write failed
This test writes and reads a 55h and AAh checker board
pattern to each address of ram U11. Any incorrect readback will cause a test failure.
This error is only readable from the remote interface.
603
A/D sync stuck
The main controller issues an A/D sync pulse to U17 and U16 to
latch the value in the ADC slope counters. A failure is detected when a sync interrupt
is not recognized and subsequent time-out occurs.
604
A/D slope convergence failed
The input amplifier is configured to the measure zero
(MZ) state in the 10 V range. This test checks whether the ADC integrator produces
nominally the same number of positive and negative slope decisions (±10%) during
a 20 ms interval.
605
Cannot calibrate rundown gain
This test checks the nominal gain between
integrating ADC and the U17 on-chip ADC. This error is reported if the procedure
can not run to completion due to a hardware failure.
Содержание E3646A
Страница 2: ...2...
Страница 16: ...16 Contents Contents...
Страница 17: ...1 Quick Start...
Страница 27: ...2 General Information...
Страница 39: ...3 Front Panel Operation and Features...
Страница 72: ...Chapter 3 Front Panel Operation and Features Calibration Overview 72...
Страница 73: ...4 Remote Interface Reference...
Страница 125: ...5 Error Messages...
Страница 136: ...Chapter 5 Error Messages Calibration Errors 136...
Страница 137: ...6 Application Programs...
Страница 148: ...Chapter 6 Application Programs Example Program for Excel 97 148...
Страница 149: ...7 Tutorial...
Страница 160: ...Chapter 7 Tutorial Remote Programming 160...
Страница 161: ...8 Specifications...
Страница 168: ...Chapter 8 Specifications Supplemental Characteristics 168 Figure 8 2 Dimensions for Rack mounting...
Страница 169: ...Appendix Service Information...
Страница 178: ...Appendix Service Information General Disassembly 178 General Disassembly...
Страница 210: ......