Chapter 5
93
Troubleshooting
Troubleshooting Gateway Configuration
4
Verify the BOOTP server is on the same local subnet as the
Gateway. If not, consult your Network Administrator to verify that
your network configuration supports using a BOOTP server that is
not on the same subnet.
5
When the Gateway is powered on, verify there is a fast flash on the
LAN Conn
LED.
Troubleshooting BOOTP with TFTP Configuration
If you have trouble configuring the LAN/GPIB Gateway using the BOOTP
with TFTP configuration method:
1
Follow the steps in “Troubleshooting BOOTP Configuration”.
2
Verify that the TFTP server system is configured to run the
tftpd
daemon. See the
Administering ARPA Services
manual for more
information.
3
Consult your BOOTP server documentation to verify you have the
correct permissions on the specified TFTP configuration file.
4
Verify that the Gateway’s configuration values are correct in the
TFTP file.
5
Look for errors specific to the TFTP values in the syslog file on the
configured syslog server or with the Telnet
syslog-display
command.
NOTE
Your BOOTP server may support the
bootpquery
capability to verify
your BOOTP configuration. See your BOOTP server documentation for
how to use the
bootpquery
command.
Содержание E2050
Страница 1: ...Agilent E2050 LAN GPIB Gateway Installation and Configuration Guide ...
Страница 2: ......
Страница 10: ...10 ...
Страница 11: ...11 1 Introduction ...
Страница 23: ...23 2 Installation ...
Страница 32: ...32 Chapter 2 Installation Installing the Hardware Notes ...
Страница 33: ...33 3 Configuration ...
Страница 63: ...63 4 Administration ...
Страница 80: ...80 Chapter 4 Administration Configuring the Gateway Notes ...
Страница 81: ...81 5 Troubleshooting ...
Страница 99: ...99 A Specifications ...
Страница 102: ...102 Specifications Specifications Notes ...
Страница 103: ...103 Glossary ...
Страница 111: ...Part Number E2050 90003 Printed in U S A E0701 E2050 90003 ...