Matrix Modules Command Reference 67
Chapter 4
Scanning Channels:
•
To scan a single channel use ROUT:SCAN (@
ssrrcc
)
•
To scan multiple channels use ROUT:SCAN (@
ssrrcc,ssrrcc
,...)
•
To scan sequential channels use ROUT:SCAN (@
ssrrcc:ssrrcc
)
•
To scan groups of sequential channels use ROUT:SCAN
(@
ssrrcc:ssrrcc,ssrrcc:ssrrcc
)
•
or any combination of the above
NOTE
Channel numbers can be in the <channel_list> in any random order.
Scanning Operation:
When a valid channel list is defined,
INITiate[:IMMediate] begins the scan and closes the first channel in the
<
channel_list
>. Successive triggers from the source specified by
TRIGger:SOURce advance the scan through the <
channel list
>. At the
end of the scan, the last trigger opens the last channel.
Stopping Scan:
See ABORt
Related Commands:
TRIGger, TRIGger:SOURce
*RST Condition:
All channels open.
Example
Scanning Using External Device
See "Scanning Channels" in Chapter 3 for examples of scanning programs
using external instruments.
Содержание E1465A
Страница 2: ......
Страница 6: ...6 ...
Страница 10: ...10 Notes ...
Страница 12: ...12 Getting Started Chapter 1 Figure 1 1 E1465A 16x16 Relay Matrix Module TERMINAL MODULE MATRIX MODULE A B C D ...
Страница 13: ...Getting Started 13 Chapter 1 Figure 1 2 E1466A 4x64 Relay Matrix Module TERMINAL MODULE MATRIX MODULE A B C D ...
Страница 14: ...14 Getting Started Chapter 1 Figure 1 3 E1467A 8x32 Relay Matrix Module TERMINAL MODULE MATRIX MODULE A B C D ...
Страница 18: ...18 Getting Started Chapter 1 Notes ...
Страница 80: ...80 Matrix Modules Command Reference Chapter 4 Notes ...
Страница 98: ...98 Register Based Programming Appendix B Notes ...