Commands by Subsystem
3
Agilent 6000 Series Oscilloscopes Programmer's Reference
75
Common (*) Commands
Commands defined by IEEE 488.2 standard that are common to all
instruments. See
"Introduction to Common (*) Commands"
Table 33 Common (*) Commands Summary
Command
Query
Options and Query Returns
*CLS (see
)
n/a
n/a
*ESE <mask> (see
*ESE? (see
<mask> ::= 0 to 255; an integer in
NR1 format:
Bit Weight Name Enables
--- ------ ---- ----------
7
128
PON
Power On
6
64
URQ
User Request
5
32
CME
Command Error
4
16
EXE
Execution Error
3
8
DDE
Dev. Dependent Error
2
4
QYE
Query Error
1
2
RQL
Request Control
0
1
OPC
Operation Complete
n/a
*ESR? (see
<status> ::= 0 to 255; an integer in
NR1 format
n/a
*IDN? (see
AGILENT TECHNOLOGIES,<model>,
<serial number>,X.XX.XX
<model> ::= the model number of the
instrument
<serial number> ::= the serial
number of the instrument
<X.XX.XX> ::= the software revision
of the instrument
n/a
*LRN? (see
<learn_string> ::= current
instrument setup as a block of data
in IEEE 488.2 # format
*OPC (see
)
*OPC? (see
ASCII "1" is placed in the output
queue when all pending device
operations have completed.
Содержание DSO6014A/L
Страница 1: ...A Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 16: ...16 Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 30: ...30 Agilent 6000 Series Oscilloscopes Programmer s Reference 1 What s New...
Страница 540: ...540 Agilent 6000 Series Oscilloscopes Programmer s Reference 6 Error Messages...
Страница 562: ...562 Agilent 6000 Series Oscilloscopes Programmer s Reference 7 Status Reporting...
Страница 584: ...584 Agilent 6000 Series Oscilloscopes Programmer s Reference 8 More About Oscilloscope Commands...
Страница 624: ...624 Agilent 6000 Series Oscilloscopes Programmer s Reference 9 Programming Examples...