:TRIGger:GLITch Commands
Table 68 :TRIGger:GLITch Commands Summary
Command
Query
Options and Query Returns
:TRIGger:GLITch:GREater
than <greater than
time>[suffix] (see
:TRIGger:GLITch:GREater
than? (see
)
<greater than time> ::=
floating-point number from 5 ns to
10 seconds in NR3 format
[suffix] ::= {s | ms | us | ns | ps}
:TRIGger:GLITch:LESStha
n <less than
time>[suffix] (see
:TRIGger:GLITch:LESStha
n? (see
<less than time> ::= floating-point
number from 5 ns to 10 seconds in
NR3 format
[suffix] ::= {s | ms | us | ns | ps}
:TRIGger:GLITch:LEVel
<level> [<source>] (see
:TRIGger:GLITch:LEVel?
(see
For internal triggers, <level> ::=
.75 x full-scale voltage from center
screen in NR3 format.
For external triggers, <level> ::= 2
volts with probe attenuation at 1:1
in NR3 format.
For digital channels (MSO models),
<level> ::= 6 V.
<source> ::= {CHANnel<n> | EXTernal}
for DSO models
<source> ::= {CHANnel<n> |
DIGital0,..,DIGital15} for MSO
models
<n> ::= 1-2 or 1-4 in NR1 format
:TRIGger:GLITch:POLarit
y <polarity> (see
:TRIGger:GLITch:POLarit
y? (see
<polarity> ::= {POSitive | NEGative}
:TRIGger:GLITch:QUALifi
er <qualifier> (see
:TRIGger:GLITch:QUALifi
er? (see
<qualifier> ::= {GREaterthan |
LESSthan | RANGe}
Содержание DSO6014A/L
Страница 1: ...A Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 16: ...16 Agilent 6000 Series Oscilloscopes Programmer s Reference...
Страница 30: ...30 Agilent 6000 Series Oscilloscopes Programmer s Reference 1 What s New...
Страница 540: ...540 Agilent 6000 Series Oscilloscopes Programmer s Reference 6 Error Messages...
Страница 562: ...562 Agilent 6000 Series Oscilloscopes Programmer s Reference 7 Status Reporting...
Страница 584: ...584 Agilent 6000 Series Oscilloscopes Programmer s Reference 8 More About Oscilloscope Commands...
Страница 624: ...624 Agilent 6000 Series Oscilloscopes Programmer s Reference 9 Programming Examples...