3 Installing BenchCel Microplate Handler
Mounting the robot
44
BenchCel Microplate Handler User Guide
Mounting the robot
About this topic
Typically, the BenchCel Microplate Handler will be installed for you. However, this topic
provides a reference for how to mount the robot on the BenchCel device.
Procedure
CAUTION
Do not use the robot arms to lift the robot. Doing so could damage the arms. Ensure
that you lift the robot by the base and the head. Dropping or bumping the robot can
damage the electronics. Handle the robot carefully.
Figure
Correct and incorrect lift points on the robot head
To mount the robot on the BenchCel device:
1
Remove the shipping bracket from the BenchCel
x
-axis carriage as follows:
a
Refer to the following figure to locate the BenchCel
x
-axis carriage.
b
Loosen the two large black bracket screws that prevent the
x
-axis carriage
from moving during shipping.
c
Use a 5-mm hex wrench to remove the four screws that secure the shipping
bracket to the
x
-axis carriage.
d
Carefully lift the shipping bracket straight up, ensuring that you do not scratch
the electronics.