A Quick reference
BenchCel Diagnostics - General Settings tab
178
BenchCel Microplate Handler User Guide
BenchCel Diagnostics - General Settings tab
You use the General Settings tab to adjust various settings. See
.
Parameters
Description
Stack Settings area
Rack sensor threshold
Any value at or higher than this means a rack is
present. Below this value means there is no rack.
Low pressure threshold
(psi)
The value at or below which the system displays a
low-pressure warning.
Default value: 30 psi
Note:
If this value is too low, the device could drop
microplates without generating a low air-pressure
warning.
Home robot after protocol
run finishes
The option to return the robot to its home position
after the protocol run is finished.
This option applies to all BenchCel devices that are
controlled by the same computer running the
VWorks software.
Speed settings
The command that opens the Speed Settings dialog
box, which enables you to set each speed (Slow,
Medium, and Fast) as a percentage of the factory-set
maximum speed.
Show current settings
The command that displays the parameter values
that were last saved.
Write settings to file
The command that allows you to save the Advanced