Software error messages
About this topic
This topic lists possible software error messages, the causes of the errors, and ways to
resolve the errors. If you are still experiencing problems with the BenchCel Microplate
Handler after trying the solutions, contact Agilent Technical Support.
Software error messages
The following software error messages might appear during a protocol run or when
you are using BenchCel Diagnostics. The messages are listed alphabetically.
Diagnosing problems
“Diagnostic tools” on page 146
Adjusting the plate-orientation sensors
“Setting sensor thresholds” on page 99
Checking and adjusting the stacker
gripper pressure (internal air pressure)
“Adjusting the stacker gripper pressure” on
page 158
Reporting problems to Agilent
Technologies
“Reporting problems” on page 162
For information about…
See…
Error message
Cause
Solution
Communication timed
out
Communication between the
controlling computer and the
BenchCel Microplate Handler or
external device failed.
Try the following:
1
Verify that the device profile has the
correct the COM port number (serial) or
Ethernet ID.
2
Restart the VWorks software.
3
Restart the BenchCel Microplate Handler.
4
Check the Ethernet or serial cables.
5
(Ethernet only) Verify that the Ethernet
switch has power.
If the problem continues, contact Agilent