7 Maintenance and troubleshooting
Hardware problems
BenchCel Microplate Handler User Guide
Hardware problems
About this topic
This topic lists possible hardware problems, the causes of the problems, and ways to
resolve the problems. If you are still experiencing problems with the BenchCel
Microplate Handler after trying the solutions, contact Agilent Technical Support.
Hardware problems
Problem
Cause
Solution
The BenchCel
Microplate Handler
does not turn on.
Your lab does not meet the
electrical requirements.
Make sure your lab meets the electrical
requirements. See
The BenchCel Microplate Handler
is not connected to the power
source.
Connect the BenchCel Microplate Handler to
the power source. See
.
The fuse is blown.
See
“Replacing the fuse” on page 134
A bad fuse could be indicative of other
problems. If the fuse blows again, contact
Agilent Technical Support.
A hissing sound can
be heard.
A leak is present in the air
connection or inside the device.
Check the air connections at the back of the
device and at the source (house, cylinder, or
pump).
If the connections look fine, the leak might be
inside the device. Contact Agilent Technical
Support.
Oil is present in the
stacker head.
The compressed air is not from an
oil-free compressor and oil has
leaked into the BenchCel
Microplate Handler.
Contact Agilent Technical Support.
The microplate drops
from the robot
grippers.
The labware definition for the
microplate type might contain
incorrect information.
Check the following in the labware definition:
Stacker gripper offset, Robot gripper offset,
Stacking thickness, Plate thickness, Sensor
intensity, and the Plate presence threshold.
To check or correct these values, open
BenchCel Diagnostics. On the
Controls
page,
click the
Labware
tab.
Note:
In the Plate offsets (mm) area, type 8 in
the Robot gripper offset and the Stacker
gripper offset boxes. The recommended
offset for most labware is 8 mm.
If the problem persists, contact Agilent
Technical Support.