![Agilent Technologies BenchBot Скачать руководство пользователя страница 47](http://html1.mh-extra.com/html/agilent-technologies/benchbot/benchbot_safety-and-installation-manual_2868485047.webp)
37
2 Laboratory setup requirements
Labware and performance specifications
BenchBot Robot Safety and Installation Guide
Labware and performance specifications
The BenchBot Robot can move all ANSI- compliant labware with or without lid.
Related information
Labware or performance parameter
Value
Distance between grippers
Maximum: 134 mm
Minimum: 77 mm
Homed: 130 mm
Payload
Maximum: 300 g
Repeatability
x, y, z
: ±0.1 mm
Yaw (xy plane): ±0.02°
Transfer time
Pick- and- place: 5 s average
For information about...
See...
General placement considerations
“General placement considerations” on
page 12
Electrical requirements
“Electrical requirements” on page 35
Environmental requirements
“Environmental requirements” on
page 36
Computer requirements
Содержание BenchBot