Aurora
Presto
—User Guide
4-22
481801
DMM Measurement Tests
You can set up a ±% variance threshold that is used to set
up the exception points on the chart.
Setting up DMM Tests
-1-
From the
Select Test Mode
screen, choose
Auto Tests
and press enter to display the
Select
Tests
screen.
-2-
Select
DMM test
and press
F2
Param
to display the
DMM Test Mode
screen.
Max DCV (±):
Enter the maximum ± DC Volts range
from 1 to 400.
Min DCV (±):
Enter the minimum ± DC Volts range
from 1 to 400.
Max Res:
Enter the maximum resistance range
from 0 to 1 MOhm.
Min Res:
Enter the minimum resistance range
from 0 to 1 MOhm.
481801-7.book Page 22 Thursday, May 23, 2002 10:46 AM
Содержание AuroraPresto
Страница 48: ...AuroraPresto User Guide 2 18 481801 481801 7 book Page 18 Thursday May 23 2002 10 46 AM...
Страница 100: ...AuroraPresto User Guide 3 52 481801 481801 7 book Page 52 Thursday May 23 2002 10 46 AM...
Страница 194: ...AuroraPresto User Guide 4 94 481801 481801 7 book Page 94 Thursday May 23 2002 10 46 AM...
Страница 220: ...AuroraPresto User Guide 7 10 481801 Approvals 481801 7 book Page 10 Thursday May 23 2002 10 46 AM...
Страница 224: ...AuroraPresto User Guide 7 14 481801 481801 7 book Page 14 Thursday May 23 2002 10 46 AM...
Страница 225: ...481801 Issue 7 05 02 Appendix 1 AuroraPresto Menus 481801 7 book Page 1 Thursday May 23 2002 10 46 AM...
Страница 233: ...481801 Issue 7 05 02 Appendix 2 ADSL Status Messages Vendor Codes 481801 7 book Page 1 Thursday May 23 2002 10 46 AM...
Страница 243: ...Issue 7 05 02 A3 1 Appendix 3 Glossary Abbreviations 481801 7 book Page 1 Thursday May 23 2002 10 46 AM...
Страница 257: ...481801 Issue 7 05 02 Index 481801 7 book Page 1 Thursday May 23 2002 10 46 AM...
Страница 266: ...AuroraPresto User Guide Index 10 481801 481801 7 book Page 10 Thursday May 23 2002 10 46 AM...