Verifying the Orbital Shaking Station configuration
AssayMAP Bravo Platform Installation Guide
Verifying the Orbital Shaking Station configuration
IMPORTANT
If the Orbital Shaking Station is installed in the AssayMAP
Bravo Platform, use the following procedures to configure and test the
accessory in every profile.
Verifying the Orbital Shaking Station configuration in a profile
To verify the Orbital Shaking Station configuration in a profile:
1
In
Bravo Diagnostics
, ensure the correct profile is selected in the
Profiles
tab.
2
In the
Configuration
tab, verify that the deck image shows a Orbital Shaking
Station icon at deck location 9.
3
To verify the settings, right- click deck location 9 in the image. In the
shortcut menu that appears, click
Configure location > Accessory
. The Bravo
Accessories Wizard opens.
4
In the
Location for accessory
list, verify the location and then click
Next
.
5
In the
Accessory already configured
page, click
Yes, reconfigure existing accessory
,
and then click
Next
.