23
4 Miscellaneous errors and recovery techniques
Pausing and continuing a run in the workbench
AssayMAP Bravo Platform Error Recovery Guide
Pausing and continuing a run in the workbench
You can pause a run to introduce labware that was forgotten during the setup, add
reagents to a reservoir that is at risk of depleting, refill or empty the wash station
carboys, and so forth.
To pause and then continue a run:
1
In the workbench application or utility navigation panel, click
Pause
.
The task currently in progress finishes before the protocol pauses. The Scheduler
Paused dialog box opens.
Note:
Finishing the task could take a long time depending on the task, for
example, the Sample Load step. If you are not sure if the task has finished, you
can use the Toggle Full Screen button to display the VWorks window and then
check the Main Log.
2
While the Bravo device is idle, you may access the Bravo deck to make changes for
your run, for example, adjusting a labware position, removing a lid, or adjusting the
volume in a reservoir.
3
To resume the run, click
Continue
in the
Scheduler Paused
dialog box.