Multimeter Command Reference
85
CONFigure
The
CONFigure
command subsystem configures the multimeter to perform the
specified measurement with the given range and resolution.
CONFigure
does not
make the measurement after setting the configuration. Executing
CONFigure
is
equivalent to setting the multimeter configuration as follows:
After configuring the multimeter, use the
INITiate
command to place the multimeter
in the wait-for-trigger state and store readings in the multimeter's internal memory.
Or, use the
READ?
command to make the measurement and send the readings to the
output buffer when the trigger is received.
Subsystem Syntax
CONFigure
:CURRent:AC [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:CURRent[:DC] [<
range
>|MIN|MAX|DEF|AUTO[,<
r
esolution
>|MIN|MAX|DEF]]
:FREQuency [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:FRESistance [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:PERiod [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:RESistance [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
:VOLTage:AC [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
[:VOLTage[:DC]] [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
[:VOLTage[:DC]]:RATio [<
range
>|MIN|MAX|DEF|AUTO[,<
resolution
>|MIN|MAX|DEF]]
The
CONFigure
command
RANGe
and
RESolution
parameters are optional. You
will get the default range and resolution settings if you do not specify a range or
resolution in the command. You will get these default settings even if you set a range
Command
Setting
RANGe
As specified (or AUTO).
RESolution
As specified, or as a function of range, integration
time, or NPLCs.
AC filter
([SENSe:]DET:BAND)
20 Hz - 300 kHz (medium filter)
Autozero
([SENSe:]ZERO:AUTO)
OFF if resolution setting results in NPLC <1;
ON if resolution setting results in NPLC
1
Input resistance
([SENSe:]INP:IMP:AUTO)
Applies to dc voltage and is disabled for all other
functions. 10M
for all dc voltage ranges.
Samples per trigger
(SAMP:COUN)
Trigger count
(TRIG:COUN)
Trigger delay
(TRIG:DEL)
Trigger source
(TRIG:SOUR)
VM Complete routing
(OUTP:TTLT<
n
>:STAT)
1 sample
1 trigger
AUTO (Automatic delay)
IMM (trigger signal is always true)
OFF (all trigger lines;
n
= 0 - 7)
Math function
(CALCulate:STATe)
OFF
Содержание Agilent E1412A
Страница 2: ......
Страница 10: ...8 Contents Notes...
Страница 14: ...12 Notes...
Страница 15: ...13 Notes...
Страница 16: ...14 Notes...
Страница 20: ...18 Agilent E1412A Multimeter Module Setup Chapter 1 Figure 1 2 Multimeter Measurement Terminals...
Страница 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Страница 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Страница 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...