Multimeter Command Reference
141
RESistance:RANGe
[SENSe:]RESistance:RANGe <
range
>
selects the range for 2-wire resistance
measurements.
Parameters
Comments
•
To select a standard measurement range, specify range as the input signal’s
maximum expected resistance. The multimeter then selects the correct range.
•
MIN
selects the minimum range available with the
RESistance:RANGe
command: 100
.
MAX
selects the maximum range available: 100M
.
•
You must select a range using
RESistance:RANGe
before specifying
resolution. Also, in order to specify an aperture time of 10ms, you must select
a fixed range.
•
Specifying a fixed range disables the autorange mode set by the
RES:RANG:AUTO
command.
•
The
RES:RANG
command overrides the range setting from a previous
CONFigure
command on the same function. The multimeter uses the same
aperture time to set the resolution on the new range as was selected by
CONFigure
.
•
*RST
Condition:
RES:RANG 1k
Example
Change the Range
CONF:RES 1320,MAX
Function: 2-wire ohms;
range selected: 10k
; MAX resolution: 1
.
RES:RANG 220
Range selected: 1k
; MAX resolution: 0.1
.
READ?
Place multimeter in wait-for-trigger state and
make measurements; send readings to output
buffer.
enter statement
Enter readings into computer.
RESistance:RANGe?
[SENSe:]RESistance:RANGe? [MIN|MAX]
returns one of the following numbers
to the output buffer:
•
The present resistance range selected if
MIN
or
MAX
is not specified. Only the
ranges available with the
RANGe
command are returned. For example, if
CONFigure
selects the 900
range, 1k
is the range returned.
•
The minimum resistance range available (100
) if
MIN
is specified.
•
The maximum resistance range available (100M
) if
MAX
is specified.
Parameter Name
Parameter Type
Range of Values
Default Units
<
range
>
numeric
100
|1k
|10k
|100k
|1M
|10M
|100M
|MIN|MAX
ohms
Содержание Agilent E1412A
Страница 2: ......
Страница 10: ...8 Contents Notes...
Страница 14: ...12 Notes...
Страница 15: ...13 Notes...
Страница 16: ...14 Notes...
Страница 20: ...18 Agilent E1412A Multimeter Module Setup Chapter 1 Figure 1 2 Multimeter Measurement Terminals...
Страница 26: ...24 Agilent E1412A Multimeter Module Setup Chapter 1 Notes...
Страница 194: ...192 Agilent E1412A Multimeter Error Messages Appendix B Notes...
Страница 201: ...Index 199 Index Agilent E1412A User s Manual and SCPI Programming Guide...