![Agilent Technologies Agilent 86120B Скачать руководство пользователя страница 178](http://html.mh-extra.com/html/agilent-technologies/agilent-86120b/agilent-86120b_user-manual_2867999178.webp)
5-46
Programming Commands
CALCulate3 Subsystem
DATA?
Queries the data resulting from delta, drift, and signal-to-noise measurements.
Syntax
:CALCulate3:DATA? {POWer | FREQuency | WAVelength | WNUMber}
Argument
Description
POWer
Queries the array of laser-line powers after the calculation
is completed.
FREQuency
Queries the array of laser-line frequencies after the
calculation is completed.
WAVelength
Queries the array of laser-line wavelengths after the
calculation is completed.
WNUMber
Queries the array of laser-line wave numbers after the
calculation is completed.
Attribute
Summary
Preset
State: not affected
SCPI Compliance: standard
Query Only
Description
The data returned by the query depends upon which calculation state is on. If
no calculation state is on, an error is generated. The returned data is comma
delimited. The following string is a typical example of six values returned
when POWer is specified from a delta power measurement:
-7.42000,-1.00000,-2.52000,
-3.41000,-3.80000,-6.36000
Notice that only measurement values are returned to the computer. There is
no first value that indicates the number of values contained in the string as
there is, for example, with the FETCh, READ, and MEASure commands. Use
the CALC3:POIN? query to determine the number of points the CALC3:DATA?
query returns.
In the SNR or ASNR calculation, only the POWer argument is valid. The other
arguments will generate a “Settings conflict” error. Use the CALC2:DATA?
query to retrieve the signal wavelengths and powers.
Содержание Agilent 86120B
Страница 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Страница 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Страница 12: ......
Страница 38: ......
Страница 68: ......
Страница 100: ...4 18 Programming Monitoring the Instrument ...
Страница 132: ......
Страница 162: ...5 30 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268 ...
Страница 230: ......
Страница 231: ...7 Definition of Terms 7 3 Specifications 7 6 Regulatory Information 7 10 Specifications and Regulatory Information ...
Страница 241: ...7 11 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 242: ...7 12 Specifications and Regulatory Information Regulatory Information Front view of instrument Rear view of instrument ...
Страница 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Страница 250: ...8 8 Reference Menu Maps System Setup Menu ...
Страница 268: ......