5-5
Programming Commands
Common Commands
*ESR?
The *ESR (event status register) query returns the value of the event status
register.
Syntax
*ESR?
Description
When you read the standard event status register, the value returned is the
total of the bit weights of all of the bits that are set to one at the time you read
the byte. The following table shows each bit in the event status register and its
bit weight. The register is cleared when it is read.
Query Response
<integer>
ranges from 0 to 255.
Table 5-3. Standard Event Status Register
Bit
Bit Weight
Condition
7
128
PON – Power On
6
64
Not Used
5
32
CME – Command Error
4
16
EXE – Execution Error
3
8
DDE – Device Dependent Error
2
4
QYE – Query Error
1
2
Not Used
0
1
OPC – Operation Complete
Содержание Agilent 86120B
Страница 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Страница 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Страница 12: ......
Страница 38: ......
Страница 68: ......
Страница 100: ...4 18 Programming Monitoring the Instrument ...
Страница 132: ......
Страница 162: ...5 30 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268 ...
Страница 230: ......
Страница 231: ...7 Definition of Terms 7 3 Specifications 7 6 Regulatory Information 7 10 Specifications and Regulatory Information ...
Страница 241: ...7 11 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 242: ...7 12 Specifications and Regulatory Information Regulatory Information Front view of instrument Rear view of instrument ...
Страница 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Страница 250: ...8 8 Reference Menu Maps System Setup Menu ...
Страница 268: ......