4-20
Programming
Monitoring the Instrument
The contents of the Standard Event Status Register can be read and the regis-
ter cleared by sending the *ESR? query. The value returned is the total bit
weights of all of the bits that are set at the present time.
Enabling register bits with masks
Several masks are available which you can use to enable or disable individual
bits in each register. For example, you can disable the Hardcopy bit in the
OPERation Status Register so that even though it goes high, it can never set
the summary bit in the status byte high.
Use the *SRE common command to set or query the mask for the Status Byte
Register.
The masks for the OPERation Status and QUEStionable Status registers are
set and queried using the STATus subsystem’s ENABle commands.
Use the *ESE common command to set or query the mask for the Standard
Event Status Register.
Table 4-4. Bits in Questionable Status Register
Bit
Definition
0, 1, and 2
not used
3
POWer
- indicating that the instrument is measuring too high of a power.
4 through 8
not used
9
Maximum signals
- indicating that the instrument has found the maximum
number of signals.
10
Drift Reference
- indicating that the number of reference signals is different
from the current number of input signals.
11
Delta Reference
- indicating that there is no delta reference signal.
12 through 13
not used
14
Command Warning
- indicating that the instrument has received some extra
unexpected parameters for one of the measurement functions.
15
not used
Содержание Agilent 86120B
Страница 1: ...Agilent 86120B Multi Wavelength Meter User s Guide ...
Страница 11: ...Contents 3 Contents Power Cords 8 16 Agilent Technologies Service Offices 8 18 ...
Страница 12: ......
Страница 38: ......
Страница 68: ......
Страница 100: ...4 18 Programming Monitoring the Instrument ...
Страница 132: ......
Страница 162: ...5 30 Programming Commands CALCulate1 Subsystem Query Response For normal update 34123 For fast update 4268 ...
Страница 230: ......
Страница 231: ...7 Definition of Terms 7 3 Specifications 7 6 Regulatory Information 7 10 Specifications and Regulatory Information ...
Страница 241: ...7 11 Specifications and Regulatory Information Regulatory Information Declaration of Conformity ...
Страница 242: ...7 12 Specifications and Regulatory Information Regulatory Information Front view of instrument Rear view of instrument ...
Страница 248: ...8 6 Reference Menu Maps Delta On Menu Delta Off Menu ...
Страница 250: ...8 8 Reference Menu Maps System Setup Menu ...
Страница 268: ......