Tunable Lasers
How to Perform a Wavelength Sweep
118
Agilent 8163A Lightwave Multimeter, Agilent 8164A Lightwave Measurement
System, & Agilent 8166A Lightwave Multichannel System, Fourth Edition
a
when the sweep is completed, or
b
when you press [Stop].
How to Perform a Manual Sweep
You can perform a manual sweep if you choose the <
Manual
> sweep mode or if
you press [Pause] during an automatic sweep. In a manual sweep you choose
when you want to perform each step of a stepped sweep. You can choose to move
forward or backward a wavelength or to end the sweep.
To perform a manual wavelength sweep over the range 1510 nm to 1570 nm,
three times, sweeping two ways, in 1 nm steps:
1
Move to the Tunable Laser channel and press [Details].
2
Move to [
Sweep
Mode
] and press
Enter.
3
Move to the <
Manual
>, by using the cursor key, and press
Enter
.
4
Perform step 4 to step 13 on page 114 to set the sweep parameters.
5
Press the [Active] hardkey on the front panel of your Tunable Laser module to
enable the optical output.
6
Press [Run Swp] to start the sweep. The screen in Figure 76 then appears.
7
You can:
–
press [Next] to move on to the next wavelength step,
–
press [Prev] to move on to the previous wavelength step, or
–
press [Stop] to end the sweep.