![Agilent Technologies Agilent 7890 Series Скачать руководство пользователя страница 341](http://html.mh-extra.com/html/agilent-technologies/agilent-7890-series/agilent-7890-series_maintenance-manual_2867996341.webp)
Maintaining a Valve
19
Maintaining Your GC
341
To Replace a Rotary Valve in the Valve Box
1
Gather the following:
•
•
T- 10 Torx screwdriver
•
1/4- inch wrench
•
Needle- nosed pliers
•
Vacuum
2
Load the
and wait for the GC to
become ready.
3
Set all valves to
Off
.
4
Leave on the GC and valve actuator air.
5
Turn off the carrier gas and sample line flows and relieve
any back pressure to the valve.
WA R N I N G
Do not install a liquid sampling valve (LSV) in the valve box if you
plan to heat the box above 75 °C. Heating an LSV over 75 °C can
cause a leak and subsequent explosion. LSVs should be mounted
in the side location to avoid potential explosions.
WA R N I N G
The oven, inlet, detector, and valve box may be very hot.
Sample and/or harmful gases may be present. Refer to your
company’s standard operating procedures for purging the
chemicals from the sample line.
Содержание Agilent 7890 Series
Страница 1: ...Agilent Technologies Agilent 7890 Series Gas Chromatograph Maintaining Your GC ...
Страница 28: ...28 Maintaining Your GC 3 Maintaining the GC ...
Страница 74: ...74 Maintaining Your GC 6 Maintaining the Purged Packed Inlet 1 2 mm ...
Страница 100: ...100 Maintaining Your GC 6 Maintaining the Purged Packed Inlet ...
Страница 126: ...126 Maintaining Your GC 7 Maintaining the COC Inlet ...
Страница 150: ...150 Maintaining Your GC 8 Maintaining the MMI ...
Страница 176: ...176 Maintaining Your GC 9 Maintaining the PTV Inlet ...
Страница 186: ...186 Maintaining Your GC 10 Maintaining the VI ...
Страница 188: ...188 Maintaining Your GC 10 Maintaining the VI 7 Install the interface See To Install the VI Interface on page 189 ...
Страница 194: ...194 Maintaining Your GC 10 Maintaining the VI ...
Страница 208: ...208 Maintaining Your GC 11 Maintaining the FID ...
Страница 258: ...258 Maintaining Your GC 13 Maintaining the uECD ...
Страница 271: ...Maintaining the NPD 14 Maintaining Your GC 271 ...
Страница 288: ...288 Maintaining Your GC 14 Maintaining the NPD ...
Страница 306: ...306 Maintaining Your GC 15 Maintaining the FPD ...
Страница 319: ...Maintaining the G3435A G3436A FPD 16 Maintaining Your GC 319 8 Lift the cover off the detector ...
Страница 328: ...328 Maintaining Your GC 17 Maintaining the Aux EPC ...
Страница 334: ...334 Maintaining Your GC 18 Maintaining the PCM ...
Страница 348: ...348 Maintaining Your GC 19 Maintaining a Valve ...
Страница 353: ...Swagelok Connections 20 Maintaining Your GC 353 Figure 14 Completed fitting ...